Sterintech Bowie Dick Helix Test (Stainless Steel PCD)
The Bowie Dick Helix Test (Stainless Steel PCD) is an ISO 11140-1 Type 2 process control test developed to verify the sterilization performance of hollow and porous loads in steam sterilizers. With its stainless steel Helix body and 4-metre test tube, steam penetration, vacuum performance and air removal are evaluated under the most challenging conditions. By helping detect leaks, poor steam quality, non-condensable gases and temperature- or time-related process errors, it supports the safe release of sterile loads.
Technical Information
The Bowie Dick Helix Test (Stainless Steel PCD) is an ISO 11140-1 Type 2 process control test developed to verify the sterilization performance of hollow and porous loads in steam sterilizers. With its stainless steel Helix body and 4-metre test tube, steam penetration, vacuum performance and air removal are evaluated under the most challenging conditions. By helping detect leaks, poor steam quality, non-condensable gases and temperature- or time-related process errors, it supports the safe release of sterile loads.
Product Features
- Compliant with ISO 11140-1:2014 Type 2.
- Suitable for checking hollow and porous loads.
- Used with a stainless steel Helix PCD and 4-metre test tube.
- Verifies steam penetration, air removal and vacuum performance.
- Helps detect steam leaks, poor steam quality, non-condensable gases and temperature- or time-related errors.
- Results are easy to interpret and document.
- Free from toxic substances and lead (Lead Free).
- The Metal PCD has a service life of more than 5,000 uses when the O-ring is replaced every 500 cycles.
Technical Information
| Product Code | Product | Pack Contents |
|---|---|---|
| 101.203.0100 | Stainless Steel Helix + 4 m Tube | 100 indicators |
| 101.203.0250 | Stainless Steel Helix + 4 m Tube | 250 indicators |
| 101.203.0500 | Replacement Indicator Set for 101.203.xxxx | 500 indicators |
Areas of Use
- Hospitals
- Central Sterile Services Departments (CSSD)
- Operating Theatres
- Dental clinics
- Veterinary clinics
- Medical device reprocessing centres
- Daily Bowie Dick and Helix performance checks of steam sterilizers
- Routine process verification of hollow and porous loads