Sterintech Hydrogen Peroxide Plasma Biological Indicator (20 Minutes)

Biological indicators are designed to check quickly whether sterilization has been effective. They consist of a plastic capsule and a vented cap that sits on top; the cap lets the sterilant in, while a filter paper in between prevents contamination. Inside the capsule is a crushable ampoule whose growth medium changes colour when bacteria grow. The bottom of the capsule contains Geobacillus stearothermophilus spores. When used with an automatic reader at 56–60 °C, it gives results with 20-minute, 1-hour, 3-hour and 24-hour options. The outside of the capsule also carries a Type 1 chemical indicator strip.


Technical Information

Biological indicators are designed to check quickly whether sterilization has been effective. They consist of a plastic capsule and a vented cap that sits on top; the cap lets the sterilant in, while a filter paper in between prevents contamination. Inside the capsule is a crushable ampoule whose growth medium changes colour when bacteria grow. The bottom of the capsule contains Geobacillus stearothermophilus spores. When used with an automatic reader at 56–60 °C, it gives results with 20-minute, 1-hour, 3-hour and 24-hour options. The outside of the capsule also carries a Type 1 chemical indicator strip.

 

PLASMA 20 MIN

  • Process: Plasma or Vaporised Hydrogen Peroxide (H₂O₂)
  • Test organism: Geobacillus stearothermophilus
  • Spore Population: 10⁶ spores / carrier
  • Label Colour Change: Initial: green → Final: yellow
  • Medium Colours: Positive: yellow; Negative: purple
  • Shelf Life: 24 months
  • ISO 11138